Magnetic Switch Guide

OpenNerve Gen2 April 2026


Overview

The Gen2 IPG includes a Hall-effect magnetic switch (MAG_DET, PA3) that allows non-contact user interaction through the implant body. It is the primary way to wake the device, put it to sleep, and reset BLE pairing credentials without any physical connection.

All interactions mediated through the MCU are gesture-based: the firmware measures how long the magnet is held and only acts when the magnet is removed, based on that duration. No action is taken on approach alone.

In addition to state changes mediated through the MCU there is an analog circuit that physically connects or disconnects the battery from the rest of the board using the magnetic switch, allowing power cycling in the event of a firmware issue or shutdown for storage. Turnon is mediated by capacitor C800, which triggers a connection between batteries and the rest of the circuit after holding the magnet for 10-15 seconds. Shutdown is mediated by capacitor C802, which disconnects power after 20-30 seconds of holding the magnet. Startup / shutdown occurs as soon as the capacitor charges to the necessary voltage level, and is not gesture-based (does not require removing the magnet).

Important note on magnetic shutdown: currently, there is an issue in the magnetic switch circuit where, once the device enters shutdown state, capacitor C802 must discharge fully before the IPG can be turned back on. Due to the high resistance between RC2_VC and GND this can take up to five minutes. Furthermore, attempting to turn on the IPG early by applying the magnet will re-charge C802 and further lengthen the discharge time. To avoid issues, 1) remove the magnet as soon as the device enter shutdown, and 2) always wait 5 minutes after shutdown before attempting to turn the IPG back on.


Magnetic Switch Circuit

Screenshot 2026-04-13 at 11 30 36 AM


Digital State Changes (sleep-wake, BLE reset)

How Duration Is Measured (Digital)

When a magnet approaches the device, a 1-second LPTIM timer starts counting. Each whole second that passes increments an internal counter. When the magnet is removed, the counter value (in whole seconds elapsed) is passed to the gesture logic.

Consequence: duration has 1-second resolution. A hold that ends at 2.9 seconds counts as 2, not 3. To reliably register a 2-second hold, remove the magnet after hearing/feeling 2 full seconds have passed — not before.

The maximum measurable duration is 255 seconds (hardware limit of the 8-bit counter). Any hold longer than ~4 minutes is treated the same as 255 seconds.


Gesture Windows

There are two independent gesture windows, both configurable over BLE by an Admin-authenticated session:

Parameter ID Default Range Purpose
Wakeup min time HP22 2 s 1–60 s Start of wake/sleep window
Wakeup max time HP23 6 s 1–60 s End of wake/sleep window
Reset min time HP24 9 s 1–60 s Start of BLE reset window
Reset max time HP25 13 s 1–60 s End of BLE reset window

A hold duration that falls within the wakeup window triggers a sleep/wake toggle. A hold that falls within the reset window counts toward a BLE credential reset. Durations outside both windows are silently ignored.

The two windows do not overlap by default (wakeup: 2–6 s, reset: 9–13 s). They can be configured to overlap, but overlapping windows mean a single gesture would trigger both actions simultaneously — generally undesirable.


Gesture 1: Sleep / Wake Toggle

Behavior

Device state when magnet removed Result
Sleeping (STATE_SLEEP) Wakes → enters BLE advertising mode
Any active state Stops all active operations and goes to sleep
DVT mode Ignored — magnet has no effect
Shutdown state Ignored — magnet has no effect

Active states that the magnet will put to sleep

The sleep transition fires from any active state including: - BLE advertising (STATE_ACT_MODE_BLE_ACT) - BLE connected (STATE_ACT_MODE_BLE_CONN) - Therapy session running (STATE_ACT_MODE_THERAPY_SESSION) - Impedance test (STATE_ACT_MODE_IMPED_TEST) - Battery test (STATE_ACT_MODE_BATT_TEST) - Wireless charging — WPT High or WPT Paused (STATE_ACT_MODE_WPT_HIGH/PAUSED)

In all of these cases, active operations (stimulation, sensing, charging) stop cleanly before sleep entry. An EVENT_MAGNET_DETECTION log entry is written whenever the gesture is recognized.

Timing with default settings

Hold duration Result
< 2 s Ignored
2–6 s Sleep ↔ wake toggle
7–8 s Ignored (gap between windows)
9–13 s Counts toward BLE reset gesture (see below)
> 13 s Ignored

Gesture 2: BLE Credential Reset

This gesture allows the device to temporarily accept connections from any BLE programmer, regardless of the stored passkey or whitelist. It is used to re-pair with a new or replacement programmer if the original is unavailable.

How it works

  1. Hold the magnet for the reset window duration (default 9–13 s), then remove.
  2. An internal counter increments. On the first qualifying hold, a 10-minute countdown window opens.
  3. Repeat the gesture until the counter reaches 5 within that 10-minute window.
  4. On the 5th qualifying hold: the device enters BLE default mode for 10 minutes.

BLE default mode

While active, app_func_ble_is_default() returns true. This causes: - The stored passkey to be replaced by the factory default passkey for the duration - The whitelist to be disabled — any programmer can attempt authentication

BLE default mode expires after 10 minutes, or immediately if the counter window expires before 5 holds are completed.

Reset gesture state machine

Event Effect
1st qualifying hold Counter = 1; 10-minute window opens
2nd–4th qualifying holds (within window) Counter increments
5th qualifying hold (within window) BLE default mode active for 10 minutes
Window expires before 5th hold Counter resets to 0; window closes
BLE default mode timer expires Returns to normal BLE credentials

The window and count are reset any time the 10-minute timer runs out without reaching 5.


EOS / Post-Shutdown Behavior

If the device has reached End of Service (EOS) and entered sleep with BATT_SW_EN pulled low (batteries in the process of disconnecting), applying the magnet with a valid wakeup gesture will still wake the device. BATT_SW_EN is restored HIGH on wakeup, which halts the battery-disconnect process. Normal BLE advertising begins. The EOS counter remains set — the next battery test (or the next return to sleep) will re-evaluate whether EOS still applies.


Summary of All Hold Durations (Default Settings)

Duration Action
< 2 s Nothing
2–6 s Wake from sleep or go to sleep from any active state
7–8 s Nothing
9–13 s Increment BLE reset counter (must repeat 5× within 10 min)
> 13 s Nothing

Relevant Source Files

File Role
Core/Src/gpio.c Configures MAG_DET_Pin (PA3) as rising+falling edge EXTI
Core/Src/stm32u5xx_it.c EXTI3_IRQHandler dispatches to HAL GPIO callbacks
App/Bsp/Src/bsp_magnet.c Rising/falling edge callbacks; LPTIM-based duration measurement
App/Functions/Src/app_func_state_machine.c app_func_sm_magnet_lost_cb() — sleep/wake state transitions
App/Functions/Src/app_func_ble.c app_func_ble_magnet_lost_cb() — BLE reset gesture counter
App/Functions/Src/app_func_parameter.c Default values for all four time window parameters